Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared
(NDIR) absorption, ratiometric measurement.
Initialization Time
Capnogram within 15 seconds, full specifications within 60 seconds.
Calibration
No calibration required.
CAPNOSTAT® Mainstream CO2
Sensor
| Size: | 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.). |
| Weight: | 18 grams (0.64 oz.), cable excluded |
| Shock Resistance: | Withstands repeated six foot drops to a hard surface. |
CO2 (Carbon Dioxide)
| Range: | 0 - 150 mmHg |
| Accuracy: | ±2 mmHg (for 0 - 40 mmHg), ±5% of reading (for 41 - 70 mmHg), ±8% of reading (for 71 - 150 mmHg) |
| Resolution: | 1 mmHg |
| Response Time: | 60 ms |
| Units: | mmHg, kPa, or % |
| Compensations: | N2O & O2 (selectable), Pb (automatic) |
| Capnogram: | Selectable sweep speed |
Respiratory Rate
| Range: | 0-150 br/min |
| Accuracy: | ±1 br/min |
| Resolution: | 1 br/min |
Sampling Flow Rate (Model 715 only)
180 ml/min (for non-intubated patients)
Response time is less than 200 ms at 180 ml/min flow
Principle of Operation
Red/Infrared absorption.
SpO2 (Oxygen Saturation)
| Range: | 0-100% |
| Accuracy: | ±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2 |
| Resolution: | 1% |
| Averaging: | Menu selectable 2 or 8 seconds |
Pulse Rate
| Range: | 30-250 bpm |
| Accuracy: | ±1% of full scale |
| Resolution: | 1 bpm |
| Averaging: | 8 seconds |
Sensors
Reusable Y-Sensor (can be sterilized and used with all patient populations);
reusable adult finger sensor
Alerts
| Limits: | Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout. |
| Audio: | Adjustable volume, two minute silence or off (LED indicators) |
| Visual: | On-screen and LED |
Display
| Type: | Dot matrix, backlit (LCD) |
| Size: | 2.2 × 1.1 inches |
Trending Capabilities
| Memory: | 24 hr., battery backed |
| Format: | 30 minute on-screen or printed in trend or tabular mode. Download data to personal computer for use with NovaCARD software |
Communication Output
RS232 to Seiko DPU-414 printer, NovaCARD software or computer
Physical
| Size (height, width, depth): | 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.). |
| Weight: | 793 grams (28 oz.). |
Electrical
External Supply (optional)
| Input: | 120 VAC 60 Hz, or 220 VAC 50 Hz |
| Battery: |
|
| Battery: | Standard AA lithium: 5.5 hour life |
Environmental
| Operating temperature: | 0 to +40 °C (32-104 °F) |
| Operating humidity: | 0 to 90% RH non-condensing. |