Specifications
TIDAL WAVE® Capnograph

Capnograph

Principle of Operation
Solid state mainstream sensor. Single beam, non-dispersive infrared (NDIR) absorption, ratiometric measurement.

Initialization Time
Capnogram within 15 seconds, full specifications within 60 seconds.

Calibration
No calibration required.

CAPNOSTAT® Mainstream CO2 Sensor

Size: 3.30 × 4.24 × 2.16 cm (1.30 × 1.67 × 0.85 in.).
Weight: 18 grams (0.64 oz.), cable excluded
Shock Resistance: Withstands repeated six foot drops to a hard surface.

CO2 (Carbon Dioxide)

Range: 0 - 150 mmHg
Accuracy: ±2 mmHg (for 0 - 40 mmHg),
±5% of reading (for 41 - 70 mmHg),
±8% of reading (for 71 - 150 mmHg)
Resolution: 1 mmHg
Response Time: 60 ms
Units: mmHg, kPa, or %
Compensations: N2O & O2 (selectable), Pb (automatic)
Capnogram: Selectable sweep speed

Respiratory Rate

Range: 0-150 br/min
Accuracy: ±1 br/min
Resolution: 1 br/min

Sampling Flow Rate (Model 715 only)
180 ml/min (for non-intubated patients) Response time is less than 200 ms at 180 ml/min flow

Pulse Oximeter

Principle of Operation
Red/Infrared absorption.

SpO2 (Oxygen Saturation)

Range: 0-100%
Accuracy: ±2% SpO2 (for 80-100% SpO2)
(1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: Menu selectable 2 or 8 seconds

Pulse Rate

Range: 30-250 bpm
Accuracy: ±1% of full scale
Resolution: 1 bpm
Averaging: 8 seconds

Sensors
Reusable Y-Sensor (can be sterilized and used with all patient populations);
reusable adult finger sensor

General

Alerts

Limits: Adjustable limits for ETCO2, SpO2, Respiratory Rate and Pulse Rate. No Respiration Detected adjustable timeout.
Audio: Adjustable volume, two minute silence or off (LED indicators)
Visual: On-screen and LED

Display

Type: Dot matrix, backlit (LCD)
Size: 2.2 × 1.1 inches

Trending Capabilities

Memory: 24 hr., battery backed
Format: 30 minute on-screen or printed in trend or tabular mode.
Download data to personal computer for use with NovaCARD software

Communication Output
RS232 to Seiko DPU-414 printer, NovaCARD software or computer

Physical

Size (height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.).
Weight: 793 grams (28 oz.).

Electrical
External Supply (optional)

Input: 120 VAC 60 Hz, or 220 VAC 50 Hz
Battery:
Nickel-Metal-Hydride (NiMH): 4.5 hour life
Battery: Standard AA lithium: 5.5 hour life

Environmental

Operating temperature: 0 to +40 °C (32-104 °F)
Operating humidity: 0 to 90% RH non-condensing.



Top Of Page